800-801-9934 Affiliates Alfa's Profile Become an Authorized Service in your area Home International Sterilizer repairs Local service near you Nail salon sterilizers Parts Replace your sterilizer Site Map Support Technical Help Testimonials Ultrasonic Cleaner What Sterilizer shall I get? Disclaimer My Account

 

Alfa Medical
14 Bond St
Great Neck, NY 11021
Contact us



EHS Series
 
 

The R.O. system contains the following 5 stages:

1. Stage 1 is a 20 Micron Particular Filter that removes dirt, sediment, sand, and other physical particles.
2. Stage 2 is a 5 Micron Particular Filter to remove small particles such as dirt, sediment, sand and other physical particles.
3. Stage 3 is Activated Carbon which reduces chemicals, such as chlorine, that may be in your water supply.
4. Stage 4 is the Reverse Osmosis Membrance which reduces 93%-98% of dissolved mineral salts and other containates in your potable water supply.
5. Stage 5 is a Post Carbone Filter which performs a polishing process that remove undesirable tastes and odors from the wate

 

Site Preparation for installation of the Riverse Osmosis System


Source Water Supply
Water Source chlorinated or non-chlorinated
Water Temp. Range 40-100F
Water Pressure Range 30-100psi
Allowable pH Range 3.0-11.0
Max. Total Dissolved Solids <2000 ppm
Turbidity <1.0 Net Turbidity (NTU)

Chemical Parameters of Source Water
Chlorine (Cl2) 0.00 ppm
Hydrogen Sulfide (H2S) 0.00 ppm
Max. Iron (Fe) <0.1 ppm
Max. Manganese (Mn) <0.05ppm
Hardness (CaCO3) <350 ppm
Bacteria water source must be potable

" Mounting of the filter unit: 2-1/4" horizontally slotted holes, 10 5/8" on center,
13" up from the bottom of the filter cartridges

Dimension:
a) Filter unit 8"W x 17"L x 18"H
b) Storage Tank 10.3 gal, 16" dia x 18"H

Weight: Filter unit 15 lbs (dry)

Tap Water Supply: The unit is supplied with 6' of ¼" O.D. x 11/64" I.D. poly tubing. A saddle valve (not supplied) is required to complete the connection to the cold water line.

Output Line: The unit is supplied with 6' of 3/8" O.D. x ¼" I.D. poly tubing. Using a saddle valve to supply tape water to this R.O. system eliminates the need for an additional shut off valve behind the EHS, as noted above. In which case an adapter 3/8" compression x ½" NPT (male) would be needed to connect the output poly tube to the hose coming off the EHS.

Drain line: The unit is supplied with 6' of ¼" O.D. x 11/64" I.D. poly tubing. The drain line can be routed into the same drain as noted above for the EHS unit or a drain outlet saddle (not supplied) can be used to connect the drain line to a suitable waste line.

Storage tank: A 10.3 gal tank with a floor base is supplied. A ¼" NPT x 3/8" O.D. 90 degree elbow fitting is required (not supplied) to connect the tank to the poly tube from the filter unit, when the tank and the filter system are located at the same approximate height. If the filter system is located higher than the tank a straight fitting should be used. The unit is supplied with 6' of 3/8" O.D. x ¼" I.D. poly tubing.

Maximum distance: from the tank to the EHS is 15 feet. Longer distances will result in a reduced flow rate.

One R.O. system is required per autoclave